Support Surface Depth (C) - Support surface should be approximately 533 mm
(21 in) deep. Allow at least 51 mm (2 in) clearance behind the sterilizer for air
circulation.
Distance To Side Wall (D)
- If the sterilizer is located next to side walls, there
should be no less than 51 mm (2 in) clearance between the side of the sterilizer
and the wall.
Distance Above Sterilizer (E)
- If the sterilizer is to be located beneath wall cabi-
nets or shelves, the underside of the cabinets or shelves should be at least 584
mm (23 in) above the support surface, to provide access to the top inspection
cover and to allow for changing the printer paper roll.
Overhang (F)
- If the sterilizer is to be located beneath an overhang, the under-
side of the overhang should project no further than 381 mm (15 in) over the rear
of the sterilizer.
Location On Support Surface
- The front of the sterilizer should be located near
the front of the support surface so water can be easily drained from the tube into
a container.
Neighboring Materials and Equipment
- If the sterilizer will be operated in con-
tinuous cycles, locate the sterilizer where excessive emission of steam will not
damage materials or equipment in the surrounding area.
Re-Location Requirements For Sterilizer
1. Disconnect power cord from the electrical outlet and allow sterilizer to cool.
2. Drain water from reservoir or take care not to tip the sterilizer, which will
allow water to spill from reservoir.