19
Standard Cycle Parameters
The following table lists the standard cycle parameters for the four different
types of sterilizer loads (follow state and federal guidelines regarding steriliza-
tion of single use devices and instruments).
List of Authorized Accessories
Listed below are the accessories which are authorized for use with this steriliz-
er:
Accessory Name
Order Number
• Printer Accessory 9A155001
• Speedclean, 1 case (12 bottles) 002-0396-01
• Speedclean, 1 bottle 002-0396-00
Standard Cycle Parameters
CYCLE TEMP / PRES-
SURE/TIME (mini-
mums)
ITEMS TO BE STERILIZED
(Always consult the item manufacturer’s recom-
mendations for sterilization.)
Unwrapped 132°C (270°F) /
186 kPa (27 psi)
for 3 minutes
Instruments loose on a tray. Open glass
or metal canisters. Tubing not used in
surgical procedures. Items manufactur-
ers recommend for exposure at 132°C
(270°F) for 3 minutes. The sterility of un-
wrapped items is compromised on expo-
sure to a non-sterile environment.
Pouches 132°C (270°F) /
186 kPa (27 psi)
for 15 minutes
Instruments and dental handpieces.
Loosely wrapped individual instru-
ments. Multiple layers of instruments
separated by fabric. Wrapped trays of
loose instruments. Tubing not used in
surgical procedures. Items manufactur-
ers recommend for exposure at 132°C
(270°F) for 15 minutes.
Liquids 121°C (250°F) /
104 kPa (15 psi)
for 30 minutes
Liquids or gels that could boil over with
rapid depressurization. Cycle includes
special vent and dry functions.
Not rec-
ommended for sterilization of liquids
intended for direct patient contact.
Packs 121°C (250°F) /
104 kPa (15 psi)
for 30 minutes
Textiles and surgical packs wrapped for
sterilization. Items, except liquids, manu-
facturers recommend for exposure at
121°C (250°F) for 30 minutes.
Operation
Return To Table Of Contents